Preliminary results of an attempt to predict over apron occupational exposure of cardiologists from cardiac fluoroscopy procedures based on DAP (dose area product) values
- 155 Downloads
Abstract
This study is an effort to propose a mathematical relation between the occupational exposure measured by a dosimeter worn on a lead apron in the chest region of a cardiologist and the dose area product (DAP) recorded by a meter attached to the X-ray tube. We aimed to determine factors by which DAP values attributed to patient exposure could be converted to the over-apron entrance surface air kerma incurred by cardiologists during an angiographic procedure. A Rando phantom representing a patient was exposed by an X-ray tube from 77 pre-defined directions. DAP value for each exposure angle was recorded. Cardiologist exposure was measured by a Radcal ionization chamber 10X5-180 positioned on a second phantom representing the physician. The exposure conversion factor was determined as the quotient of over apron exposure by DAP value. To verify the validity of this method, the over-apron exposure of a cardiologist was measured using the ionization chamber while performing coronary angiography procedures on 45 patients weighing on average 75 ± 5 kg. DAP values for the corresponding procedures were also obtained. Conversion factors obtained from phantom exposure were applied to the patient DAP values to calculate physician exposure. Mathematical analysis of our results leads us to conclude that a linear relationship exists between two sets of data: (a) cardiologist exposure measured directly by Radcal & DAP values recorded by the X-ray machine system (R 2 = 0.88), (b) specialist measured and estimated exposure derived from DAP values (R 2 = 0.91). The results demonstrate that cardiologist occupational exposure can be derived from patient data accurately.
Keywords
Interventional cardiology Entrance surface air kerma (ESAK) Over apron exposure Cardiologist DAPIntroduction
Interventional cardiology includes a number of procedures such as coronary angiography (CA), percutaneous transluminal coronary angioplasty (PTCA) and electrophysiology procedures that are widely used in routine clinical practice. CA is used for detection of cardiovascular diseases. This procedure consists of injection of a contrast medium into the blood vessels of the patient through a catheter and highlighting the anatomical region of interest using fluoroscopy with X-rays [1, 2, 3].
During these procedures patients are exposed to significant doses due to the long duration of fluoroscopy and the large numbers of cine frames which are taken in each examination. Furthermore, the nature of interventional fluoroscopy examinations requires the presence of physicians (cardiologists) in the close vicinity of the patient for relatively long times. The result is much higher occupational dose compared to other diagnostic procedures [4, 5]. The ever increasing number of cardiovascular patients worldwide requires a careful programme for protection of patients as well as personnel of the related departments [4, 6].
Scattered radiation arising from the patient is the main contributor to the dose to medical staff. Leakage radiation from the X-ray tube and, in some cases, the primary beam itself are the other possible sources of exposure [3, 7, 8]. Therefore, occupational doses in catheterization laboratories (CATHLABs) may exceed limits recommended by the International Commission on Radiological Protection (ICRP), and may even induce deterministic radiation effects in cases where radiological protection tools are not used appropriately [8, 9]. Interventional cardiology is a clinical practice with a high radiation risk, therefore appropriate protection evaluations should be performed, occupational doses should be monitored carefully and special attention should be paid to radiation protection of the staff involved in these procedures [4, 10, 11, 12].
To reduce occupational dose various protection tools such as leaded aprons, goggles, bed side curtains and glass screens suspended from the ceiling over the patient’s body are used [11]. It is also necessary to implement effective measures for accurate, fast and simple dosimetry appropriate for such environments.
Limits for occupational exposure are expressed in effective dose [8]. Effective dose is intended for estimation of the overall stochastic risk from exposure to ionizing radiation. This quantity is calculated as the weighted sum of organ doses [8, 13]. For monitoring of the dose received by radiation workers the personal dose equivalent (H p(10)) is measured. H p(10) is an estimator of the absorbed dose at 10 mm depth in human tissue. When a worker’s body is uniformly exposed by radiation, the effective dose (E) can be estimated simply from H p(10) [14, 15, 16]. However, exposures in cardiology and interventional suites are typically non-uniform.
Presently, various methods are used to estimate effective dose arising from different fluoroscopic techniques [3, 7, 17]. Different methods have been proposed to derive effective dose for radiation workers from doses measured at one or two points on their body surface [16, 18, 19]. For example, in Britain, based on ICRP 85 recommendations, it is recommended to place one dosimeter under the lead apron to measure H p(10). By contrast, in the United States of America, according to NCRP 122, it is advised that if only one dosimeter is used, it must be placed over the shield at the neck region and if two dosimeters are worn, one should be placed under the apron at the lumbar region and the other over the apron at the neck region [11]. Recently, one radiation protection center in Holland (NCR) has proposed an alternative protocol. In their report, they recommended the use of only one dosimeter over the lead apron to increase the measurement’s accuracy and to lower the costs. This technique will typically result in higher reported dose compared with the actual dose value. This method may be preferred to the models which underestimate the dose and consequently the radiation risks [11, 20]. In summary it should be noticed that there are costs involved in application of these practically used methods and the obtained results are typically retrospective.
To estimate the exposure of the patient, dose area product (DAP) can be utilized. ICRP has reported that the exposure of the medical practitioner correlates with that of the patient [4, 21, 22]. In recent years, several research teams have tried to find what relationship may exist between the patient’s recorded DAP values and cardiologists’ dose received during cardiac fluoroscopy. The main goal of these studies were to utilize DAP values to estimate physician dose arising from interventional fluoroscopy. Bor et al. 2009 have investigated the accuracy of two dosimetry techniques to estimate the doses to cardiologists. The first method included establishment of a database to relate the cardiologist extremity doses to the DAPs of patients. In the second one, an equation was suggested from which the cardiologist doses were calculated. Finally, they concluded that, as a useful method, DAP values may be used to estimate the cardiologists’ doses [5]. Kuipers et al. 2010 demonstrated the relationship between absorbed dose measured over the lead apron in the chest region of the cardiologists and the patient exposure (DAP). They found a linear relationship (correlation coefficient of R 2 = 0.55) between these two variables [21] though this correlation is not very strong. In another study carried out by the European DIAMOND cardiology research group in 2006, the relationship between cardiologists’ dose and their patients’ exposure (DAP) was also investigated [23]. These previous studies examined correlations between total DAP and measures of occupational dose, however they took no account of the different imaging projections used. Better correlations may be expected when the amount of radiation delivered by each of the contributing imaging projections is considered.
Materials and methods
DAP to Exposure conversion coefficients arising from exposing a Rando phantom
The main goal of this study was to obtain the factors that may be used to estimate a cardiologist entrance surface air kerma (ESAK) arising from a CA or PTCA procedure.
Measured exposure and recorded DAP value for a specific view
Locations of the dosimeter, physician phantom, patient phantom, X-ray tube, and the patient table
Schematic diagrams illustrating the positions of the cardiologist and patient phantoms
Definition of imaging views to produce converting factors
Different reference views in LAO and RAO directions of coronary fluoroscopy X-ray tube to image right coronary artery (RCA) and left coronary artery (LCA)
| Hurst’s the heart [2] | Braunwald’s heart disease [1] |
|---|---|
| RAO 30° | LAO 60°, Cranial 20° |
| LAO 45° | LAO 60°, Caudal 25° |
| LAO 45°, Cranial 30° | R/LAO 0°, Caudal 20° |
| RAO 15°, Cranial 30° | R/LAO 0°, Cranial 20° |
| LAO 45° | RAO 30°, Cranial 20° |
| LAO 45°, Cranial 30° | RAO 30°, Caudal 25° |
| LAO 60° | |
| LAO 60°, Cranial 25° | |
| RAO 30° |
X-ray tube positions in RAO, LAO directions (part A); cranial (CR), caudal (CA) (part B) directions are illustrated. Each view is shown in a specific range with light red lines
DAP to ESAK converting factors (CF); calculations and applications
An example of the informatics files generated by the AXIOM Artis system at the end of each cardiac fluoroscopy procedure
Cardiologist exposure measurement in the presence of a real patient
To provide a comparison with the estimation from the DAP; the ESAK at the chest region of the cardiologist was measured by the Radcal ionization chamber for 45 cases. Patient weight ranged from 70 to 80 kg. The cardiologist was recommended to avoid any unnecessary movement. A suspended leaded glass curtain was also used to reduce scattered photons impinging physician’s body. For each procedure the informatics file of the Siemens AXIOM Artis system was extracted and the patient’s DAP values were broken down into DAP values for various projections (Fig. 4). At the end of each procedure two ESAK values were made available: measured and estimated exposure. The measured exposure was obtained as explained earlier. The estimated exposure was reconstructed from potentially 77 different scenarios and one fluoroscopy mode; as described in “DAP to ESAK converting factors (CF); calculations and applications” section. Finally physician’s measured and estimated exposures are compared. If the two set of ESAK values are not significantly different, it implies that the proposed method may be useful to estimate cardiologist’s exposure indirectly, simply, but accurately.
Results
Measured cinegraphy and conversion factors from DAP to physician ESAK (nGy/µGym2) for 77 specified imaging views for an AXIOM Artis fluoroscopy system
| Cinegraphy Views | RAO 90° | RAO 80° | RAO 60° | RAO 40° | RAO 20° | R/LAO 0° | LAO 20° | LAO 40° | LAO 60° | LAO 80° | LAO 90° |
|---|---|---|---|---|---|---|---|---|---|---|---|
| CR 40° | 5.95 | 3.86 | 2.48 | 1.52 | 1.03 | 1.32 | 6.35 | 6.44 | 8.92 | 9.43 | 10.71 |
| CR 25° | 5.91 | 2.96 | 2.02 | 1.37 | 1.09 | 1.28 | 4.74 | 5.20 | 7.50 | 7.55 | 10.18 |
| CR 10° | 5.37 | 2.17 | 2.08 | 1.29 | 1.06 | 1.25 | 3.24 | 4.37 | 6.45 | 8.46 | 7.51 |
| CR0°, CA0° | 4.91 | 2.18 | 1.92 | 0.97 | 0.83 | 0.94 | 3.08 | 4.18 | 5.20 | 9.07 | 7.28 |
| CA 10° | 5.65 | 1.88 | 1.53 | 1.12 | 1.01 | 0.93 | 2.02 | 4.06 | 6.66 | 7.45 | 9.48 |
| CA 25° | 5.56 | 2.99 | 1.72 | 1.07 | 0.88 | 0.87 | 1.97 | 3.79 | 5.26 | 6.49 | 8.12 |
| CA 40° | 5.19 | 3.00 | 1.58 | 1.18 | 0.88 | 0.91 | 1.71 | 2.94 | 4.32 | 5.70 | 7.32 |
Mean, standard deviation, first, second (median) and third quartiles of measured over-apron ESAK, corresponding estimated ESAK, recorded DAP values and percentage overestimation of ESAK arising from 45 angiographic procedures
| Values | Mean | SD | 25th | 50th | 75th | Minimum | Maximum | Normality (KS test) |
|---|---|---|---|---|---|---|---|---|
| Measured ESAK (nGy) | 3,143 | 1,960 | 1,573 | 2,950 | 4,530 | 630 | 11,587 | 0.61 |
| Estimated ESAK (nGy) | 3,625 | 2,160 | 1,764 | 3,449 | 5,257 | 788 | 10,541 | 0.31 |
| Total DAP (µGym2) | 1,216 | 736 | 576 | 1,078 | 1,747 | 269 | 3,392 | 0.09 |
| Percentage overestimation of ESAK | 14.5 | 17.2 | 5.0 | 18.0 | 26.5 | −20.0 | 48.0 |
Box plot of occupational exposure values of the cardiologist (nGy) which are presented for the two sets of data: a measured exposure and b estimated exposure; the black lines within the boxes mark the median, the box demarks the 25th and 75th percentile (the interquartile range) and the whiskers represent the highest and lowest values that are not outliers (values that are more than 1.5 times the interquartile range)
The plot shows; (a and b) the relationship between the measured and estimated exposure with the recorded DAPtotal (µGym2) values in each angiography procedure, respectively
The plot shows the relationship between measured and estimated exposure values in each CA procedure
Discussion
Converting factors were obtained for 77 tube positions covering the total space around the patient’s body. As expected the factor is highest for the larger angle RAO or LAO views, and also the larger Cranial (CR 40°) and Caudal (CA 40°) angulations. This is due to the increased volume of tissue in the path of the primary X-ray beam producing more scattered radiation. Also, when the tube is advancing towards the CR 40o view it moves beyond the leaded table drape. In these circumstances the shorter distance to the physician dosimeter would result in higher contribution from scattered radiation. It should be noted that the leakage from the tube also contributes to the measured exposure but to a lesser extent compared to the scattered radiation.
Average exposure measured and estimated in this study (Table 3) are very close to the results reported by Häusler et al. in 2009 (4 µGy with 0–14 µGy range) [29]. Our results for average and range of total DAP values per procedure are comparable with the corresponding values obtained by Kuon et al. [30] and Vano et al. [10]. The results of Kuon and Vano et al. are (12.9 Gycm2 with 3–30 Gycm2) respectively. The marginal differences between our results and those reported by other researchers are in favor for the accuracy of our results. In other words, factors which influence physician exposure have been correctly chosen. These factors include patient weight, procedure type, protective material and the method of selection and use of the protective material. In this study we have tried to reduce the effect of confounding factors on the physician dose, thus the correlation between DAP values and cardiologist exposure is improved considerably (R 2 = 0.88), when compared with the corresponding figures reported by Tsapaki et al. (R 2 = 0.43) [23] and Kuipers et al. (R 2 = 0.55) [21].
Patient weight and size have significant effects on the cardiologist exposure; a large patient produces more scattered radiation from his/her body [4, 9]. Kuipers et al. did not include patients of limited weight in their study, while in this work patients within the weight range of 75 ± 5 kg have been selected. Kuipers et al. presented results of both CA & PTCA procedures while our data are acquired only from CA examinations. Cardiologist exposures during CA and PTCA procedures are different from each other. In PTCA, due to the longer period of patient exposure, the physician exposure is higher [7, 23]. Kuipers et al. included seven physicians in their work; our data have all been obtained from examinations performed by one cardiologist. In other words, each physician (depending on his/her skills and experience) behaves differently in identical conditions [18]. This could be one of the main reasons for the differences observed between two sets of results in the aforementioned studies.
The results obtained in the present study cannot be simply and generally extended, because the different sizes of patients, variety of operators and the type of the procedure will increase the variability of the results and the related uncertainties. However, a comparison between our results and the others gives an indication of the uncertainty that is normally encountered in description of these variables.
Correlation between DAP and estimated ESAK for the physician would be 100 % if only a single conversion factor was used. The use of more conversion factors for more narrowly defined projections would be expected to decrease the correlation between total DAP and estimated ESAK. A strong correlation between total recorded DAP and physician estimated exposure (R 2 = 0.95) was observed in this work, in spite of the division of the rotational space into relatively small angles (77 views). Additionally, a larger number of tube positions should provide a more accurate estimation of the cardiologist ESAK due to the use of various DAP components in comparison to simply use of a compound DAP. For this reason, various DAPs for different small angles and different modes of exposure (fluoroscopy and cine) were extracted separately for each patient. The existence of a strong correlation between measured and estimated physician exposure (R 2 = 0.91) achieved in this work is evidence that the conversion factors obtained in this study can be employed to estimate cardiologist exposure with reasonable accuracy.
Conclusions
This study is the first step towards calculating of physician’s over apron ESAK arising from CA procedures in a fast and accurate manner. However further studies are required to allow for variation of factors such as patient’s weight which have been assumed constant in this work. In a more comprehensive work PTCA procedure has to be included.
Notes
Acknowledgments
This article is based on part of the results extracted from a thesis code no: A-358 presented for the degree of Master of Science to medical physics department of Mashhad University of Medical Sciences. The authors would like to thank Javad-Al-Aeme Heart Hospital (Mashhad, Iran) for their sincere cooperation. This work was fully funded by the office of vice president for research affairs of Mashhad University of Medical Sciences.
References
- 1.Libby P, Bonow RO, Zipes DP, Douglas LM (2008) Braunwald’s Heart disease: a textbook of cardiovascular medicine. Saunders, PhiladelphiaGoogle Scholar
- 2.Fuster V, Walsh RA, O’Rourke RA, Poole-Wilson P (2007) Hurst’s the heart. McGraw-Hill Professional, New YorkGoogle Scholar
- 3.Ferrari P, Venturi G, Gualdrini G, Rossi PL, Mariselli M, Zannoli R (2010) Evaluation of the dose to the patient and medical staff in interventional cardiology employing computational models. Radiat Prot Dosim 141:82–85CrossRefGoogle Scholar
- 4.International Commission on Radiological Protection. (2013) Radiological protection in cardiology. ICRP Publication 120. Ann ICRP 42(1)Google Scholar
- 5.Bor D, Olgar T, Onal E, Caglan A, Toklu T (2009) Assessment of radiation doses to cardiologists during interventional examinations. Med Phys 36:3730–3736CrossRefPubMedGoogle Scholar
- 6.Vaño E, Gonzalez L, Fernandez JM, Alfonso F, Macaya C (2006) Occupational radiation doses in interventional cardiology: a 15-year follow-up. Br J Radiol 79:383–388CrossRefPubMedGoogle Scholar
- 7.Morrish OW, Goldstone KE (2008) An investigation into patient and staff doses from X-ray angiography during coronary interventional procedures. Br J Radiol 81:35–45CrossRefPubMedGoogle Scholar
- 8.International Commission on Radiological Protection (2007) Recommendations of the international commission on radiological protection. ICRP Publication 103. Ann ICRP 37(2–4):2Google Scholar
- 9.Vano E, Gonzalez L, Fernandez JM, Prieto C, Guibelalde E (2006) Influence of thickness and operation modes on occupational and patient radiation doses in interventional cardiology. Radiat Prot Dosim 118:325–330CrossRefGoogle Scholar
- 10.Vano E, Ubeda C, Leyton F, Miranda P, Gonzalez L (2009) Staff radiation doses in interventional cardiology: correlation with patient exposure. Pediatr Cardiol 30:409–413CrossRefPubMedGoogle Scholar
- 11.Martin CJ (2009) A review of radiology staff doses and dose monitoring requirements. Radiat Prot Dosim 136:140–157CrossRefGoogle Scholar
- 12.Vano E (2003) Radiation exposure to cardiologists: how it could be reduced. Heart 89:1123–1124CrossRefPubMedCentralPubMedGoogle Scholar
- 13.Technical reports series. Dosimetry in diagnostic radiology: an international code of practice. TRS No. 457. Vienna: International Atomic Energy Agency (2007)Google Scholar
- 14.The International Commission on Radiation Units and Measurements. Quantities and Units in Radiation Protection Dosimetry. ICRU Report 51. Bethesda, MD (1993)Google Scholar
- 15.Martin CJ, Sutton DG (2002) Practical Radiation Protection in Health Care. OUP OxfordGoogle Scholar
- 16.Clerinx P, Buls N, Bosmans H, de Mey J (2008) Double-dosimetry algorithm for workers in interventional radiology. Radiat Prot Dosim 129:321–327CrossRefGoogle Scholar
- 17.Prlic I, Suric Mihic M, Vucic Z (2008) Active electronic personal dosemeter in interventional radiology. Radiat Prot Dosim 132:308–312CrossRefGoogle Scholar
- 18.Fardid R, Toossi MTB, Mehrpouyan M, Ghorbani M (2013) Evaluation of occupational radiation exposure of cardiologists in interventional radiography in Mashhad CATHLABs. Int J Low Radiat 9:160–168CrossRefGoogle Scholar
- 19.Delichas M, Psarrakos K, Molyvda-Athanassopoulou E, Giannoglou G, Sioundas A, Hatziioannou K, Papanastassiou E (2003) Radiation exposure to cardiologists performing interventional cardiology procedures. Eur J Radiol 48:268–273CrossRefPubMedGoogle Scholar
- 20.Kuipers G, Velders XL (2009) Effective dose to staff from interventional procedures: estimations from single and double dosimetry. Radiat Prot Dosim 136:95–100CrossRefGoogle Scholar
- 21.Kuipers G, Velders XL, Piek JJ (2010) Exposure of cardiologists from interventional procedures. Radiat Prot Dosim 140:259–265CrossRefGoogle Scholar
- 22.The International Commission on Radiation Units and Measurements (2005). Patient Dosimetry for X Rays used in Medical Imaging. ICRU Report 74. Bethesda, MDGoogle Scholar
- 23.Tsapaki V, Kottou S, Vano E, Parviainen T, Padovani R, Dowling A, Molfetas M, Neofotistou V (2005) Correlation of patient and staff doses in interventional cardiology. Radiat Prot Dosim 117:26–29CrossRefGoogle Scholar
- 24.Bozkurt A, Bor D (2007) Simultaneous determination of equivalent dose to organs and tissues of the patient and of the physician in interventional radiology using the Monte Carlo method. Phys Med Biol 52:317–330CrossRefPubMedGoogle Scholar
- 25.Siiskonen T, Tapiovaara M, Kosunen A, Lehtinen M, Vartiainen E (2007) Monte Carlo simulations of occupational radiation doses in interventional radiology. Br J Radiol 80:460–468CrossRefPubMedGoogle Scholar
- 26.Järvinen H, Buls N, Clerinx P, Miljanic S, Nikodemová D, Ranogajec-Komor M, Struelens L, d’Errico F (2008) Comparison of double dosimetry algorithms for estimating the effective dose in occupational dosimetry of interventional radiology staff. Radiat Prot Dosim 131:80–86CrossRefGoogle Scholar
- 27.Bogaert E, Bacher K, Thierens H (2008) A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipment. Radiat Prot Dosim 128:312–323CrossRefGoogle Scholar
- 28.Pantos I, Patatoukas G, Katritsis DG, Efstathopoulos E (2009) Patient radiation doses in interventional cardiology procedures. Curr Cardiol Rev 5:1–11CrossRefPubMedCentralPubMedGoogle Scholar
- 29.Häusler U, Czarwinski R, Brix G (2009) Radiation exposure of medical staff from interventional x-ray procedures: a multicentre study. Eur Radiol 19:2000–2008CrossRefPubMedGoogle Scholar
- 30.Kuon E, Glaser C, Dahm JB (2003) Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures. Br J Radiol 76:406–413CrossRefPubMedGoogle Scholar